Chapter 7: Repair Procedures
Lower Device Repairs
7-48
Service Manual
6.
Remove the 14 mounting screws that secure the heat sink. Remove the heatsink.
7.
Remove the retaining clip and slide the cartridge out of the heat sink.
8.
To reassemble, seat the fan seal in the center bore of the heatsink and use it to align the fan shaft
before fully securing the heat sink with its mounting screws. It may be necessary to wet the fan hub to
allow it to slide all the way onto the fan shaft and seat properly.
7.9.4 Elevating Base
.
1.
Run the elevating base to its top travel limit.
2.
Remove the mattress, bed, translation deck, tilt platform, and pan.
3.
Run the canopy to its lowest position (closed). Use the shipping locks (Figure 5-14) to lock the canopy in
the down position.
4.
Switch off and unplug the device.
5.
Remove the chassis cover. (Refer to “Removing the Chassis Cover with the Storage Drawer in Place” on
page 7-45.)
6.
Disconnect the elevating base electrical connector at J-45 on the relay board, and tie a wire to it so you
can fish it back through during reassembly.
7.
Connect the replacement elevating base to the relay board, switch the device back on and using the
bed raise/lower pedals fully extend the new base. Switch off and unplug the device.
FIGURE 7-48. Bed Disassembly
Mattress
Clear Plate
Rotating bed
Translation deck
Tilt platform
Tilt mechanism
Pan
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...