Chapter 6: Alarms and System Failure Codes
System Failure Codes
6-26
Service Manual
6.2.42 System Failure 83
System failure 83 indicates: Unexpected SBC Software problem.
6.2.43 System Failure 85
6.2.44 System Failure 98
System failure 98 indicates: Overflow.
6.2.45 System Failure 99
System failure 99 indicates: Logic problem on the control board.
Cause(s)
Action(s)
Unexpected SBC Software
problem
Power cycle the device. If issue persists, replace the SBC.
Cause(s)
Action(s)
Defective flat ribbon cable
between control board and
SBC
Make sure ribbon cable is properly connected and if the problem is not
resolved, replace the ribbon cable.
Control board issue
Replace control board.
SBC issue
Replace SBC.
Cause(s)
Action(s)
Software on the control
board is not performing
correctly.
Replace the control board.
Cause(s)
Action(s)
Software on the control
board is not performing
correctly.
Replace the control board.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...