Chapter 7: Repair Procedures
Display Module Repairs
Service Manual
7-71
7.19.5 SBC (Single Board Computer) PCA
7.19.5.1 SBC Disassembly
1.
Remove the LCD display by following the steps in “LCD Display Assembly” on page 7-68.
2.
Working from the back of the device, disconnect the power harness from the SBC PCA by pulling firmly
outward on the white Molex connector. Disconnect the data (ribbon) cable connector by lifting the two
locking levers simultaneously. Remove the USB connector IEC plug.
3.
Disconnect the Alarm PCA Harness (LINCAN) by firmly pulling up on the connector as shown.
Note:
It may be necessary to use a plier or similar tool to remove connectors. Do not pull on the wires.
FIGURE 7-70. Disconnect Power Harness
FIGURE 7-71. Disconnect Alarm PCA Harness
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...