Chapter 8 - Replacement Procedures
8 - 7
8-2-5
Filter Replacement Procedure
8-2-5-1
Manpower
One person, 30 minutes
8-2-5-2
Tools
No special tools are needed
8-2-5-3
Filter Replacement Procedure
1.) Remove Filter Cover as described in
"Filter Cover Removal Procedure" on page 8-6
.
2.) Pull out plastic plugs and remove the filter.
Two different Plastic plugs have been used, the one to the left on early Vivid 7 scanners and the
type to the right on newer units (both Vivid 7 and Vivid 7 PRO).
3.) Install a new or a cleaned filter (GE Vingmed Part No: FB307903).
Use plastic plugs shown to the right (GE Vingmed Part No: 080X6520)
4.) Install Filter Cover as described in
"Filter Cover Installation Procedure" on page 8-6
.
Figure 8-6 Filter Cover and Filter
Figure 8-7 Plug replacement
Filter
Plastic plugs
Filter Cover
Replace this plug…
with this plug