Chapter 4 - Functional Checks
4 - 5
4-2-2
Power Shut Down
4-2-2-1
Overview
When the unit is switched off, the system performs an automatic shutdown sequence. The unit can be
switched off into two states:
•
Standby Mode:
Most of the system is powered down, but a certain portion of the unit remains energized. The
standby mode allows a shorter reboot time when the system is used on a daily basis or moved from
one place to another.
•
Full Shut Down:
The entire system is shut down. It is recommended to perform a full shutdown at least once a week.
4-2-2-2
Power Shut Down to Standby Mode
This procedure describes the needed steps to power down the unit to Standby mode:
1.) Press the
ON/OFF
key on the Control Panel once to display the System - Exit menu on the screen.
.
Figure 4-1 Press ON-OFF key once
Press this key.