Chapter 8 - Replacement Procedures
8 - 5
8-2-3
Front Cover Replacement Procedure
8-2-3-1
Manpower
One person, 30 minutes
8-2-3-2
Tools
Phillips screwdriver
8-2-3-3
Preparations
1.) Shut down the system as described in
"Power Shut Down" on page 4-5
.
2.) Disconnect all Probes and I/O cabling. See User manual.
3.) Remove Side Covers as described in
"Side Covers Removal Procedure" on page 8-3
.
8-2-3-4
Front Cover Removal Procedure
1.) Unscrew five (5) Phillips screws that fasten it to the chassis.
2.) Pull it outwards to free it from the Top Cover and Front Bumper (See
Figure 8-4
).
8-2-3-5
Front Cover Installation Procedure
1.) Thread the same, or a new, cover so it fits in between chassis and bumper (See
Figure 8-4
).
2.) Thread it so it links together with Top Cover. You may have to loosen two (2) Phillips screws on the
top cover that are closest to the Front cover.
3.) Fasten Front Cover with five (5x) Phillips screws.
4.) Fasten two (2x) Phillips screws on Top Cover if these were loosened.
5.) Install Side Covers as described in
"Side Covers Installation Procedure" on page 8-4
.
6.) New Front Cover: Install correct labels, see
"Labels Locations" on page 1-14.
Figure 8-4 Front Cover Replacement
Screws