Chapter 5 - Components and Functions (Theory)
5 - 23
Section 5-5
Patient I/O (Physio)
The Patient I/O is mounted inside the Backend Processor’s cabinet with its connector panel available
on the front of the scanner.
The Patient IO contains the electronics for ECG, phono and two analog inputs. One of the inputs has
programmable gain. The module extracts respiration from ECG. The electronics also contains a
pacemaker detector. The four inputs are separately isolated because of safety requirements.
Available inputs
•
Phono (from a phono heart microphone)
•
ECG / Respiration
•
AUX1
•
AUX2 (Pulse/Pressure)
These signals are optically isolated from the rest of the system.
Figure 5-8 Pat I/O - Block Diagram
Digital Trace Data
PHONO
AUX 1
ISOLATION
AUX 2
ECG / RESPIRATION
(Pulse/Pressure)
PHONO
ECG
AUX1
AUX2
A/D
(Pulse/Pressure)
to PC2IO