Chapter 8 - Replacement Procedures
8 - 25
8-3-10-2
Application SW Installation
1.) Insert the applicable CD labelled
Vivid7 & EchoPAC PC Application SW
, according to
Table 8-13
on page 8-25,
into the CD drive.
2.) Hold down the power switch on the operator panel for a few seconds to power down the scanner
(forced shut down).
3.) Wait until the LCD’s on the Operator Panel is switched OFF and the Power Switch on the Operator
Panel has turned yellow.
4.) Power ON the System as described in
Table 8-5 on page 8-11
.
5.) A dialogue box, called Start Application, will appear, select INSTALL SW.
6.) A dialogue box, called Startloader, will appear, select OK to continue.
7.) A dialogue box will appear, also named Startloader stating “New System SW installation will start
now”, select OK to continue.
8.) A Command prompt window will appear, hit any key on the alphanumeric keyboard to continue
9.) A message in the Command prompt window may prompt you that copying application software
takes 2 minutes.
THIS WILL TAKE APPR. 5 MINUTES.
10.)Several dialogue boxes will appear on the screen during the SW load.
11.)Remove the CD from the CD drive when prompted.
12.)The Scanner will now power down and reboot. Verify that the command prompt window is selected.
Table 8-13 Application Software List
SOFTWARE
DESCRIPTION
PART NUMBER
SW. v.2.2.0
Vivid7 & EchoPAC PC
Application SW
(BT’02-M4-2)
FC200220-04
SW. v.1.1.3
Vivid7/EchoPAC PC Application
SW
(BT’01-M4)
FC200090-06
NOTICE