Chapter 4 - Functional Checks
4 - 47
4-3-11
Measurements and Multi Image Checks
4-3-11-1
Introduction to Measurements & Analysis (M&A)
•
A complete M&A software package for Cardiac is included in the Vivid 7 / Vivid 7 PRO.
•
The M&A software runs on the Backend Processor.
•
In this test, we just check one measurement, Calipers.
4-3-11-2
Preparations, M&A
. .
4-3-11-3
Check Caliper Measurement, Distance
.
Figure 4-18 Measurement Keys
Table 4-26 Measurement Key Checks
Step
Task
Expected Result(s)
1.
Generate a 2D image
(see
4-3-4 - 2D Mode (B mode) Checks
for details).
A 2D image is active on the screen
2.
Press
FREEZE
on the Control Panel
The last scanned picture is on the screen.
3.
Scroll back until you find the picture you want to do the
measurement on.
The picture of interest is on the screen.
Table 4-27 Measurement Key Checks
Step
Task
Expected Result(s)
4.
Press
MEASURE
on the Control Panel.
Select
Caliper
in the Measurement menu.
or
Press
CALIPER
on the Control Panel.
Press the
CALIPER
assignable.
The Measurement menu with a list of studies is
displayed in the Parameters window on the screen.
When entering the Measurement mode for the first
time, the Caliper tool is selected by default.
When re-entering the Measurement mode, the first
measurement in the actual study that has not been
performed is selected by default.
Measure
Assign and Measure
Measure and Assign
Caliper
Measure and Assign
(when in measure
mode))
Menu
Fast access to the
Measurement menu
(when in measure mode)