Chapter 7 - Diagnostics/Troubleshooting
7 - 63
7-4-4-2 System Test Screen Description (cont’d)
7-4-4-2-2
Mode (Calibration/Test Screen Selection)
7-4-4-2-3
“Block Schematics” with Single Tests
Click one of the buttons in the “block schematics” in the upper part of the screen (or one of the
duplicate buttons in the lower row of the screen) to select a particular test.
Some of the tests are described after the figure.
7-4-4-2-4
Receiver Board (RX) Test
Table 7-5
Mode (Calibration/Test Screen selection)
Field
Key
Description
Mode
The buttons in this field are used to select the Calibration Screen or the Test Screen.
Calibration
Click this button to select the Calibration Screen.
Test
Click this button to select the Test Screen.
Figure 7-59 Diagnostics Test Screen, Upper Part
Table 7-6
Receiver Board (RX) Test
Test (Key) Sub-tests
Description
RX
Receiver Board (RX)
Test All
Activates the following tests, in sequence, see description for each sub test;
•
RECV0
•
RECV1
RECV.0
Max Gain test for RX using FEC signal generator.
RECV.1
Reduced Gain test for RX using FEC signal generator.
Block Schematics