GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
1 - 10
Section 1-2 - Important Conventions
1-2-3
Product Icons
The following table describes the purpose and location of safety labels and other important information
provided on the equipment.
Table 1-8
Product Icons
LABEL/SYMBOL
PURPOSE/MEANING
LOCATION
Identification and Rating Plate
Manufacturer's name and address
Date of manufacture
Model and serial numbers
Electrical ratings
Rear of console near power inlet
Type/Class Label
Used to indicate the degree of safety or
protection.
IP Code (IPX8/IP68)
Indicates the degree of protection
provided by the enclosure per IEC 529.
IPX8 indicates drip proof and may be
used in an Operating Theater.
The footswitch delivered with Vivid 7 /
Vivid 7 PRO is IP68 rated.
Footswitch
Equipment Type BF (man in the box
symbol) IEC 878-02-03 indicates B
Type equipment having a floating
applied part.
Probe connectors including Doppler
probe connector
Equipment Type CF (heart in the box
symbol) IEC 878-02-05 indicates
equipment having a floating applied part
having a degree of protection suitable
for direct cardiac contact.
ECG connector and surgical probes
Device Listing/Certification Labels
Laboratory logo or labels denoting
conformance with industry safety
standards such as UL or IEC.
Rear of console
CAUTION - This unit weighs...Special
care must be used to avoid..."
This precaution is intended to prevent
injury that may result if one person
attempt to move the unit considerable
distances or on an incline due to the
weight of the unit.
On the console where easily seen
during transport
"DANGER - Risk of explosion used in..." The system is not designed for use with
flammable anesthetic gases.
Rear of console
“CAUTION” The equilateral triangle is
usually used in combination with other
symbols to advise or warn the user.
Various