GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8 - 45
8-4-6
Front Casters Replacement Procedure
8-4-6-1
Manpower
1 persons, 30 minutes per Caster
8-4-6-2
Tools
•
Torx screwdriver T-20
•
Open end wrenches, 10 mm and 134 mm
•
5 mm hexagon key
•
Bevel Edged Board and Wooden Wedge (included in package with replacement Casters)
8-4-6-3
Preparations
1.) Shut down the system as described in
"Power Shut Down" on page 4-5
.
2.) Remove the Front Bumper as described in
"Front Bumper Removal Procedure" on page 8-50
.
3.) Turn Front Casters as shown in
Figure 8-39
.
4.) Activate Direction Lock.
5.) Put the Bevel Edged Board on the floor.
Figure 8-38 Bevel Edged Board and Wooden Wedge.
Wooden Wedge
Bevel Edged Board