5 - 40
Section 5-10 - Power Distribution
5-10-4-5
Fuses
•
None.
•
A current limiter will switch off the power if + 3.3 V, +5 V or +15 V outputs are overloaded
•
Over Voltage Protection is provided for these voltages:
-
+ 3.3 V
-
+ 5 Vd
-
+/- 5 Va
-
+/- 15 Va
•
Voltage will be turned off if the temperature grows too high.
5-10-4-6
Jumpers and Dip-switches
None.
5-10-4-7
LEDs
None.
(Observe the LEDs on Internal I/O to check DC Voltage Status, see
"Internal I/O" on page 5-79
.)