Section 5-3 - Front End Processor
5-3-7
Front End Subsystem
The Front End section performs the transmit and receive functions.
•
The Front End Controller controls the ultrasound transmitter and receiver synchronization.
•
The TX128 board makes the ultrasound transmit pulses.
•
The RX128 board is the ultrasound receiver for the ultrasound echoes from blood/tissue.
•
The two Beamformer boards digitizes the received ultrasound signal and sends it to the Mid
Processors.
•
The Relay board is the interface between the Probes and the Front End.
Figure 5-3 The Front End
RELAY
BOARD
RX128
TX128
Front End
Controller
Probes
Doppler Probe
Digitized received
ultrasound signals to
RFT Board
VME bus
XD BUS
Beam
Former 64
FE BUS