Section 3-6 - Preparing for Installation
Section 3-6
Preparing for Installation
3-6-1
Physical Inspection
•
Verify that the system arrived intact (visual inspection).
If damaged, refer to “
Damage In Transportation
“on page
v
in this manual.
•
Verify that all items are present (see the Packing List)
3-6-2
EMI Protection
This unit has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
See
Chapter 2
,
"EMI Limitations" on page 2-4
for more information about EMI protection.
Section 3-7
Completing the Installation
3-7-1
System Specifications
3-7-1-1
System Requirements Verification
•
Verify that the site meets the requirements listed in
Chapter 2
(see:
"Facility Needs" on page 2-7
)
•
Verify that the specifications below don’t conflict with any on site conditions
3-7-1-2
Physical Dimensions
The physical dimensions of the Vivid 7 / Vivid 7 PRO unit are summarized in
Table 3-5
.
3-7-1-3
Weight with Monitor and Peripherals
3-7-1-4
Acoustic Noise Output
Less than 70dB(A) according to the standard DIN 45635 - 19 - 01 - KL2.
Table 3-5
Physical Dimensions of Vivid 7 / Vivid 7 PRO with monitor and peripherals
Height
Width
Depth
Unit
137.5 - 157.5
64
90
cm
54.1 - 62
25.2
35,4
Inches
Table 3-6
Weight of Vivid 7 / Vivid 7 PRO with monitor and peripherals
Model
Weight [kg]
Weight [lbs]
Vivid 7 / Vivid 7
PRO
200
410