Chapter 4 - Functional Checks
4 - 17
4-3-4
2D Mode (B mode) Checks
4-3-4-1
Introduction
The 2D Mode is the system’s default mode.
4-3-4-2
Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner’s left-most active probe connector.
-
See
"Probe Connection" on page 3-13
for info about connecting the probes.
-
See
Chapter 9 -
, section
9-17
for available probes.
2.) Turn ON the scanner.
-
See
4-2-1 - Power ON/ Boot Up
on page
4-2
for info about booting the scanner.
The 2D Mode window is displayed (default mode).
3.) If needed, adjust the Display’s Brightness and Contrast setting (
See “Monitor Adjustments” on
page 6-2.
).
Figure 4-11 2D Mode Screen Example
Status Window
Control Menu