Section 7-2 - Diagnostics
7-2-9-4
DC Offset Calibration Procedure
1.) Disconnect all connected probes.
2.) Select
DC Offset calibration
and then click the EXECUTE button to start the DC Offset calibration
of the Beam Formers.
3.) Wait for the calibration to take place. When finished, restart Vivid 7 / Vivid 7 PRO.