Section 3-9 - Connectivity
3-9-3-7-8
Setting the DICOM Server as Default Dataflow
1.) Finnish the verification in
"Check the Connection to the DICOM Server (Ping)" on page 3-53
.
You need to have green check marks on both Hardware Ping and DICOM Ping before you continue.
1.) Click the DATAFLOW tab.
2.) Select “DICOM Server” dataflow on the “Dataflow Name” drop-down menu.
.
3.) Check the Default box.
.
4.) Exit Config and restart the scanner (by hitting CTRL-ALT-R at the same time).
Figure 3-59 Selecting DICOM Dataflow Services
Figure 3-60 Set DICOM as Default Dataflow
DATAFLOW tab
Dataflow Name
drop-down menu
Select
“DICOM Server”
Click once in this check box
to make DICOM the Default
dataflow