GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
Chapter 3 - Installation
3 - 55
3-9-3-7-9
Performing a Study
1.) After restart, hit New Exam (you’ll see that DICOM Server is the default dataflow), enter a last name
and patient ID and start scanning and image acquisition. Images are stored in the local image buffer
on the system.
2.) After having finished the study, click on New Exam and select All when asked if you want to save.
The images will now be transferred to the server via a spooler.
3.) To view the status of the spooler job, press the F4 key on the keyboard. You can select the Refresh
button to get an update of the status of the spool job.
4.) Once the jobs are transferred, they will be deleted from the spooler.
3-9-3-7-10
Going Offline
1.) If you go offline with the scanner, you can still use the DICOM Server dataflow, and images will be
saved into the spooler.
2.) Then, when connecting to the server, go into the spooler and highlight all the studies. Then select
SEND TO… and select the server name (e.g. ProSolv server) from the Send to Service and click
on SEND. All selected studies will then be sent to the server.