GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
8 - 50
Section 8-4 - FRU Replacement Procedures
8-4-7
Front Bumper Replacement Procedure
8-4-7-1
Manpower
One person, 30 minutes
8-4-7-2
Tools
•
Phillips size 2
•
Torx screwdriver T-20
8-4-7-3
Preparations
1.) Shut down the system as described in
"Power Shut Down" on page 4-5
.
2.) Remove Front Cover as described in
"Front Cover Removal Procedure" on page 8-5
.
8-4-7-4
Front Bumper Removal Procedure
1.) Unscrew 4 Torx screws under the bumper.
2.) Lift the bumper off.
8-4-7-5
Front Bumper Installation Procedure
1.) Put the new bumper in position.
2.) Fasten the 4 screws.
NOTE:
Gentle. Don't use power.
3.) Install Front Cover as described in
"Front Cover Installation Procedure" on page 8-5
.
Figure 8-44 System front and Front Bumper
Screw holes
Front Bumper