Section 8-4 - FRU Replacement Procedures
8-4-11
Power Supply Battery Pack Replacement Procedure
8-4-11-1
Tools
Phillips screwdriver 2
8-4-11-2
Manpower
One person, 1 hour
8-4-11-3
Preparations
1.) Shut down the system as described in
"Power Shut Down" on page 4-5
.
2.) Remove Left Side Cover as described in
"Side Covers Removal Procedure" on page 8-3
.
8-4-11-4
Power Supply Battery Pack Removal Procedure
1.) Remove Backend Processor Cover.
2.) Remove the four screws from Power Supply Battery Pack cover plate.
3.) Disconnect Battery Pack connection wire.
4.) Remove Power Supply Battery Pack from Backend Processor.
Figure 8-48 Power Supply Battery Pack
1
3
4
Battery Pack
Phillips screws