GE M
EDICAL
S
YSTEMS
D
IRECTION
FC091194, R
EVISION
02
V
IVID
7 / V
IVID
7 PRO S
ERVICE
M
ANUAL
4 - 28
Section 4-3 - Functional Checks
4-3-6
Color Mode Checks
4-3-6-1
Introduction
Color Flow screens is 2D or M Mode screens with colors representing blood or tissue movement.
Color Flow may be selected both from 2D mode or from M mode or a combination of these.
4-3-6-2
Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-
See
"Probe Connection" on page 3-13
for info about connecting the probes.
-
See
Chapter 9 -
, section
9-17
for available probes.
2.) Turn ON the scanner.
-
See
4-2-1 - Power ON/ Boot Up
on page
4-2
for info about booting the scanner
The 2D Mode window is displayed (default mode).
4-3-6-3
Color 2D Mode:
Press
COLOR
.
4-3-6-4
Color M Mode:
1.) Press
MM
(M Mode) to bring up an M Mode display on the screen.
2.) Press
COLOR
.