Troubleshooting
Problem
Cause
Solution
The system does not power
up while operating from
battery power.
Battery is fully discharged.
Connect the system to an AC
outlet to charge the battery
The system powers down
while operating from battery
power.
Battery is fully discharged
Connect the system to an AC
outlet to charge the battery
You are prompted to enter
User ID
and/or
Password
while attempting to export
records to a shared network
directory.
The
User ID
and/or
Password
defined on
the
Communication Setup
window (
) are
incorrect.
1.
Press
Esc
to close the
prompt.
2.
Exit the export program.
3.
Run
Communication
Setup
.
4.
Enter the correct
User
ID
and
Password
for the
shared directory and
save the new values.
5.
Export the records.
User cannot log on to the
system.
High Security Mode
is
enabled and the user’s
User
ID
or
Password
were entered
incorrectly.
Try the following:
• Verify the user is setup in
the system.
Refer to
“User Setup” on
page 161
.
• Verify the user typed the
User ID
and
Password
correctly.
• Contact the administrator
to reset the user’s
User ID
or
Password
.
• Contact GE Healthcare
technical support to
obtain a temporary
supervisor password.
2053535-002K
MAC™ 2000
191
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......