System Configuration
•
Date of Test
•
Export Date
•
Secondary ID
5.
Press
Save
and return to the
System Configuration
window.
Retrieving Your Password
If the system is set up for
High Security Mode
and you forget your password, use the
following procedure to access the system:
1.
Contact GE Healthcare Technical Support and provide the serial number of
the device you want to access.
GE Healthcare Technical Support generates a temporary, device-specific name
and password that you can use for 24 hours.
2.
Log in to the system with the user ID
MACService
and the password provided
by GE Healthcare Technical Support.
NOTE:
If the keypad on the device does not include the letters for the
MACService
user ID, type
6227378423
for the user ID.
3.
Immediately after logging into the system, verify the name and password for
your device.
4.
Record this information and store it in a secure location for future reference.
Exporting the Audit Trail
The
Audit Trail Export
function copies the system audit trail in XML format to an SD
card and then clears the audit trail on the system. If a previous audit trail exists on the
SD card, it IS overwritten automatically by the new audit trail.
GE Healthcare recommends exporting the audit trail weekly to long term storage to
meet archive requirements. If the audit trail is not exported regularly, it consumes
storage space and reduces the number of ECGs that you can store on the device
To export an audit trail, the following conditions must be met:
•
High Security Mode
must be enabled.
To enable
High Security Mode
, see
“Basic Setup” on page 113
•
Audit Trail
must be enabled.
To enable
Audit Trail
, see
“Basic Setup” on page 113
• You must have
Edit Setup
and
Delete Records
permissions set.
To set permissions for Edit Setup and Delete Records, see
“User Setup” on page 161
.
2053535-002K
MAC™ 2000
171
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......