Setting Up the Equipment
Use the following instructions to connect the system to an AC power outlet.
Item
Description
1
Female end of the device’s power cord connected to the back of the device.
2
Male end of the device’s power cord connected to an AC outlet.
1.
Connect the female end of the device’s power cord (1) to the AC power connector
on the back of the device.
2.
Plug the male end of the device’s power cord (2) into an AC outlet.
NOTE:
It is recommended that you plug the device into an uninterruptible power
supply (UPS) or a surge suppressor.
3.
Check the Power LED to make sure the device is receiving power form the AC
outlet.
Connecting the Patient Cable
This system supports a variety of patient cables.
WARNING:
HIGH-FREQUENCY BURNS — Use of cables not supplied with this equipment can
lead to serious injury.
Use only the acquisition cable that ships with this equipment.
CAUTION:
INACCURACIES IN ECG Improper connection can cause inaccuracies in the ECG.
Trace each individual leadwire from its acquisition cable label to the colored
connector and then to the proper electrode to ensure that it is matched to the
correct label location.
46
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......