Recording a Resting ECG
ECG Options–First Row (cont'd.)
Option
Description
150 Hz
.
NOTE:
The initial
measurement
displayed is
set in
System
Configuration
>
Resting ECG
Setup
.
Eliminates noise in the waveform by restricting which frequencies
are included. Frequencies are measured in Hertz (Hz) and include
the following options:
• 20 Hz
• 40 Hz
• 100 Hz
• 150 Hz
Selecting a frequency eliminates signals that exceed that
frequency. The smaller the frequency selected, the more signal is
filtered out. For example, a filter of 40 Hz displays only signals at
40 Hz or less; signals greater than 40 Hz are ignored.
Changing the measurement here does not change the
measurement set in
System Configuration
.
Pace Enhance
Improves the readability of pacemaker ECGs. Options are
On
and
Off
.
More
Toggles between the first row of options (previous) and the second
row of options (following).
ECG Options–Second Row
Option
Description
Printer Leads
NOTE:
The initial
measurement
displayed is
set in
System
Configuration
>
Resting ECG
Setup
.
Selects which leads to include in the printout. Options are:
• First 6
• Second 6
• Rhythm 6
• 12
Use this option only when conducting rhythm ECGs. For more
information, see
“Generating a Rhythm Report (Manual Recording)
” on page 81
.
Changing the measurement here does not change the
measurement set in
System Configuration
.
ADS
Toggles the anti-drift system (
ADS
)
on
and
off
. ADS helps reduce
baseline drift.
Full Disclosure
Press to start generating a Full Disclosure ECG report.
See
“Full Disclosure” on page 77
for detailed instruction.
Main Menu
Exits
Resting ECG
and returns you to the
Main Menu.
More
Toggles between the first row of options (previous) and the second
row of options (following).
74
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......