System Configuration
Arrhythmia Setup Fields–Page 3
Field
Description
Lead Sequence
Determines the lead sequence to use.
Arrhythmia
Setup
includes the following options in addition to
the four options available in the
Resting ECG Setup
:
•
STD_C
•
STD_RED
•
STD_LI
•
CABR_LI
•
NEHB_6
•
HIGH_C
1–6 Rhythm Leads
Six fields that allow you to define the rhythm leads
and their sequence. You can select the rhythm leads
for all four lead sequences.
If you chose any of the following in
Lead Sequence
,
the rhythm leads are not displayed:
• STD_C
• STD_RED
• STD_LI
• CABR_LI
• NEHB_6
• HIGH_C
Stress ECG Setup
Stress ECG Setup
is available only if the
ERGO Stress Test
option was activated. For
more information, see
“Options Setup” on page 165
.
The
Stress ECG Setup
differs from the resting or arrhythmia ECGs. In addition to
defining the stress ECG settings, you can create, edit, or delete test protocols.
Stress ECG Settings
The
Stress ECG Setup
function allows you to define:
• Waveform parameters
• Lead usage
• Analysis options
• Report options
• Lead sequence
To access the Stress ECG Setup, on the
Main Menu
press
System Configuration
>
Stress ECG Setup
.
Many of the fields on the
Stress ECG Setup
windows are the same as those on the
Resting ECG Setup
or the
Arrhythmia Setup
. The following tables list the settings that
2053535-002K
MAC™ 2000
133
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......