Stress Testing
The menu options change. Continue to step
.
10.
Do any of the following, as necessary.
• Press
Next Patient
to test another patient.
You are warned that testing another patient discards the results of the
current test. Do one of the following:
• Press
No
to cancel the change in patients and return to the current test.
You can either print the current test report or change the report formats.
• Press
Yes
to erase the current test results and test a new patient.
Repeat from step
for the next patient.
• Press
to print the test’s report.
The report prints with the selected format options.
• Press
Report Format
to modify the report format.
The
Report Format
window opens. Select the options you want to include in
the report and press
Save
. You can now print the test’s report.
Conducting a Stress Test with a Master’s Step Device
Use the following instructions to conduct a stress test with a Master’s Step device,
if it is selected in
Basic Setup
.
1.
Prepare the patient as described in
“Preparing the Patient” on page 51
2.
On the
Main Menu
press
Stress ECG
.
The
Enter Patient Data
window opens.
3.
Enter patient data as described in
“Entering Patient Information” on page 57
Be sure you enter accurate information for
Date of Birth
,
Gender
, and
Weight
.
The number of steps is determined by these three parameters.
For more information on using Master’s Step, see
“Master’s Step Data” on page
.
4.
Adjust the stress options as necessary:
This includes the speed and gain, finite residual filter, pacemaker enhancement,
printer leads, test protocol, report format, and target heart rate. For more
information on setting these options, see
“Stress Options” on page 98
.
5.
Record a preliminary ECG.
This may be seated, standing, supine, or hyperventilating, depending on the
requirements of the selected protocol.
6.
Begin the pretest phase to allow the patient to warm up.
a.
Remove the leadwires from the patient, but leave on the electrodes.
This prevents the patient from tripping on the leadwires during the test.
b.
Instruct the patient to take a step whenever the system beeps.
c.
Press the
Pretest
key.
102
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......