Creating Barcodes
Manual Bar Code Reader Configuration Fields (cont'd.)
Field
Description and Byte Length
Visit length
The length of the patient visit ID.
The value can be from 0 to 19.
First name offset
The patient's first name
Offset
.
First name length
The patient's first name
Length
.
Be aware of the following criteria when setting the length:
• value can be from 0 to 20
• should equal the length from the MUSE CV system with
which the MAC system communicates.
NOTE:
The MAC system does not support “long” names; if the
MUSE system uses long names, this field should be set
to its maximum value.
Last name offset
The patient's last name
Offset
.
Last name length
The patient's last name
Length
.
Be aware of the following criteria when setting the length:
• value can be from 0 to 40
• should equal the length from the MUSE CV system with
which the MAC system communicates
NOTE:
The MAC system does not support “long” names; if the
MUSE system uses long names, this field should be set
to its maximum value.
Year of birth offset
The year the patient was born. Enter the field's
Offset
Year of birth length
The year the patient was born. Enter the field's
Length
The length must be set to 4.
Month of birth offset
The month the patient was born. Enter the field's
Offset
.
Month of birth length
The month the patient was born. Enter the field's
Length
.
The length must be set to 2.
Day of birth offset
The day the patient was born. Enter the field's
Offset
.
Gender offset
The patient's gender. Enter the field's
Offset
.
Gender length
The patient's gender. Enter the field's
Length
.
The length must be set to 1.
Configuring the Barcode Reader Automatically
You can configure the barcode reader automatically by scanning a barcode that has
been set up using the following information:
2053535-002K
MAC™ 2000
197
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......