Troubleshooting
Refer to
“Replacing the Battery” on page 179
for instructions on installing the
battery.
• Verify the device is connected to an AC power outlet.
Refer to
“Connecting the AC Power” on page 45
for instructions.
• Verify the equipment is receiving power from the outlet.
If the device is receiving power, the
Power LED
is lit.
ECG Data Contains Noise
If the acquired ECG data displays unacceptable noise levels, do the following:
• Check the patient’s position.
The patient should remain motionless during the acquisition of a resting ECG.
• Use the
Hookup Advisor
indicator to help determine the cause of the noise.
For more information, refer to
“Hookup Advisor” on page 70
• Verify the electrodes are placed properly.
Refer to
“Electrode Placement” on page 52
for information on proper electrode
placement.
• Verify the electrodes are applied correctly.
You must remove perspiration, excessive hair, lotions, and dead skin cells from
the electrode site.
Refer to
“Preparing the Patient’s Skin” on page 51
for more information.
• Check for defective or expired electrodes.
Replace the electrodes if there are any questions about their effectiveness.
• Check for defective, broken, or disconnected leadwires.
Replace the leadwires if there are any questions about their effectiveness. Refer to
“Connecting the Patient Cable” on page 46
.
• Consider using filters,
ADS
, and
FRF
to help eliminate or reduce ECG noise.
For more information, refer to
“ECG Options” on page 73
,
, or
“Stress Options” on page 98
External Stress Equipment does not Move
If the external stress equipment does not move automatically when expected, do
the following:
• Verify the correct stress equipment is selected in
Basic Setup
.
For more information, refer to
“Basic Setup” on page 113
.
• Verify the selected stress equipment is supported.
For a list of supported stress equipment, refer to
(Stress Option)” on page 49
.
• Verify the stress equipment is connected to the cart.
External stress equipment is connected to the cart through a serial cable. For more
information, refer to
“Rear View” on page 32
• Verify the protocol is set up to activate the stress equipment.
The protocol can set the stress equipment’s speed and grade or load. For more
information, refer to
“Editing Stress Protocols” on page 137
• Verify the
Stop TM
button is not depressed.
For more information, refer to
“Stress Test Keys” on page 98
2053535-002K
MAC™ 2000
185
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......