Maintenance
Storing Thermal Paper
Refer to the instructions for use that accompany the thermal paper for information
on storing of the thermal paper.
Battery Maintenance
The device uses a rechargeable battery containing lithium-ion cells. The battery
contains an integrated electronic fuel gauge and a safety protection circuit.
Because of the bias current needed to operate the integrated electronics, the battery
discharges even when it is not installed in the device. The rate at which it discharges
is dependent on the ambient temperature at which it is stored. The higher the
temperature, the more quickly it discharges. To prolong the battery’s charge when not
in use, store the battery in a cool, dry location.
A new, fully-charged battery should last for approximately 3 hours of normal
operation. An on-screen LED indicates the condition and capacity of the battery’s
charge. (For more information on the battery gauge, refer to
“Front View” on page 32
and
“System Errors” on page 190
). When the LED flashes amber, connect the device to
AC power to charge the battery to full capacity.
As the battery ages, the full charge capacity of the battery degrades and is
permanently lost. As a result, the amount of charge that is stored and available for
use is reduced. When the capacity is no longer sufficient for your daily operation,
you need to replace the battery.
Replacing the Battery
WARNING:
ENVIRONMENTAL HAZARD — Do NOT dispose of the battery by fire or burning.
Follow local environmental guidelines concerning disposal and recycling.
Use the following procedure to replace the battery:
1.
Unplug the device from the AC adapter.
2.
Gently turn the device over and remove the screw holding the battery cover.
3.
Push the tab to remove the cover of the battery compartment.
2053535-002K
MAC™ 2000
179
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......