Recording a Resting ECG
The Hookup Advisor is displayed as a three-circle indicator in the upper right corner
of the screen.
The following table describes each of the indicator’s conditions.
Hookup Advisor Indicators
Indicator
Description
Red
Indicates a leadfail condition or extreme baseline shifts.
The red indicator is always the left-most circle of the indicator and flashes
when lit.
Yellow
Indicates muscle artifact, power line interference, baseline wander, or
electrode noise.
The yellow indicator is always the middle circle of the indicator.
NOTE:
In RR Analysis mode, the yellow indicator is not active. RR Analysis
supports only the red and green indicators of Hookup Advisor.
Green
Indicates acceptable signal quality.
The green indicator is always the right-most circle of the indicator.
When the lead quality is red or yellow, a message describing the lead problem or
status is displayed on the screen.
Hookup Advisor continuously reviews the ECG data for acceptable lead quality.
When an ECG is acquired, Hookup Advisor runs a complete and more comprehensive
assessment of the full 10 seconds of ECG data and possibly prompts the user
regarding any poor lead quality conditions.
• If
Preview before analysis
is turned off in the system setup, a lead quality message
and prompt may be displayed, depending on the current lead quality level and the
Prompt level in the system setup. If a message and prompt is displayed, the lead
quality indicator will reflect the overall 10-second lead quality.
• If
Preview before analysis
is enabled, the system setup Prompt level is disregarded
and the system immediately displays the Preview screen. Any lead quality
messages will be displayed in this screen along with the overall 10-second lead
quality indicator.
2053535-002K
MAC™ 2000
71
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......