Recording a Resting ECG
Full Disclosure Report Setup (cont'd.)
Fields
Description and Action
Report Format
Click the drop-down arrow to determine how the full disclosure
ECG report exports to a PDF file.
The options are:
• one lead @ 25mm/s
Lead
Click the drop-down arrow to determine which lead to print
on the PDF file.
The options are:
• I
• II
• III
• aVR
• aVL
• aVF
• V1
• V2
• V3
• V4
• V5
• V6
Special Considerations
When recording ECGs, you need to make special considerations for the following
situations:
• Recording ECGs of pacemaker patients
• Recording ECGs during defibrillation
Recording ECGs of Pacemaker Patients
Because of slow paper speed, pacer pulses cannot be displayed directly on the ECG
recording. For example, with a paper speed of 50 mm/s and a pulse duration of only
0.5 ms, the width of the recorded pacer pulse would be only 0.025 mm.
If
Pace Enhance
is enabled, the recorder reduces the pulse amplitude and expands
its width to make pacer pulses easier to identify. The system records the pulse with
the correct polarity, a width of 5 ms, and equal amplitude in all leads. Depending on
the polarity of the pacer pulse in leads I and II, the pacer pulse in lead III may be
80
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......