Order Manager
3.
Enter the location(s) from which you want to retrieve orders.
Locations must match the locations used on the remote system. Separate
multiple locations with commas (for example 1, 13, 55).
4.
Press
Enter
.
The system connects to the remote system and retrieves a list of matching
orders.
5.
To select one order from the list, use the
Select
function key to select the order
you want and press
Enter
.
If you need to select multiple orders, use the
trimpad
and the
Enter
key to
highlight multiple orders.
6.
After you have selected all the orders you want to download, press the
Load
Orders
function key.
The system loads and stores the selected orders.
7.
Proceed to
“Selecting and Completing Orders” on page 65
The downloaded order list displays the
Patient Name
,
Patient ID
,
Room
,
Time
,
Type
,
Location
, and
Order Number
. The list changes as you navigate the list. You can
select and execute only one order at a time from the list. When you select an order,
the resting ECG application opens and the
Patient Information
window is populated
with the patient demographics from the selected order.
An order is completed when the ECG record is saved or transmitted to the MUSE or
non-MUSE system. Completed orders are marked as completed.
Selecting and Completing Orders
After you have orders on the system, use the following procedure to select and
complete them.
1.
On the
Order Manager
window, choose
Select
.
The cursor moves to the list of available orders.
2.
Select the order you want to use and press
Enter
.
A window opens with the order details.
3.
Do one of the following:
• To select a different order, select
Cancel
.
The detail window closes and you return to the
Order Manager
window.
• To use the selected order, select
Okay
.
• If the
TIPI
option is activated, the
Patent Information
window opens with
the information from the order.
• If the
TIPI
option is not activated, select
Patient Data
to open the
Patient
Information
window.
4.
Enter or correct the patient data.
5.
Acquire an ECG for the order and save or transmit the acquired ECG.
2053535-002K
MAC™ 2000
65
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......