4
Preparing the Patient
This chapter provides the procedures for preparing the patient's skin and properly
placing electrodes.
NOTE:
These instructions do not cover the application of electrodes for the KISS Electrode
Application System (not available in the United States). To use the KISS system, see
the KISS operator’s manual for instructions.
Preparing the Patient’s Skin
Careful skin preparation is the key to an interference-free ECG. Signal quality is
indicated on the device via the Hookup Advisor indicator.
1.
Select the electrode placement sites for ECG monitoring or diagnosis per the
protocol specified by the hospital or physician.
Refer to
“Electrode Placement” on page 52
for diagrams and descriptions of
electrode placement for various protocols.
2.
Ensure that each site is dry, clean, and free of excessive hair.
NOTE:
Do not use solvents to clean the skin; solvents trapped under electrodes
may lead to abnormal skin reactions.
3.
To prepare for a stress test, do the following:
a.
Mark each electrode site with a felt tip pen.
b.
Degrease each site with a skin preparation cream.
c.
Use a mild abrasion to remove the mark left by the felt tip pen.
2053535-002K
MAC™ 2000
51
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......