System Configuration
Resting ECG Setup Fields–Page 2 (cont'd.)
Field
Comment
QTC Calculation
Determines which formula is used to correct QT
calculations. Available options are:
• Bazett
QTc = QT √HR\60
Bazett is available only if the MEHR or MIHR option
is activated.
• Framingham
QTc = QT + 154 (1 — 60/HR)
Framingham is available only if the ME12 or MI12
option is activated.
• Fridericia
QTc = QT
3
√HR/60
Fridericia is available only if the ME12 or MI12
option is activated.
NOTE:
In all formulas, HR = Heart Rate.
Screening Criteria
Enables/disables the inclusion of the screen criteria.
This setting is available only if the
MI12
option is
activated.
If this filed is enabled,
Suppress reason statements
field is available.
Suppress normal statement
Enable/disables the inclusion of the normal
statement.
This setting is available only if the
MI12
option is
activated.
Suppress abnormal/borderline
Enable/disables the inclusion of the
abnormal/borderline statements.
This setting is available only if the
MI12
option is
activated.
Suppress all statements
Enable/disables the inclusion of all statements.
This setting is available only if the
MI12
or
MIHR
option is activated.
Suppress reason statements
Enable/disables the inclusion of reason statements.
This setting is available if the
MI12
option is activated.
122
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......