Maintenance
4.
Gently lift the cover of the battery compartment.
5.
Remove the battery from the compartment.
6.
Place the new battery in the compartment and push until it clicks into place.
7.
Replace the cover on the battery compartment.
It should click into place.
8.
Tighten the screw to hold the cover in place.
Conditioning the Battery Pack
To maintain the storage capacity of the battery installed in the device, GE
Healthcare recommends that you condition the battery once every 6 months to
recalibrate its electronic fuel gauge. A condition cycle consists of an uninterrupted
“charge-discharge-charge” cycle.
Use the following instructions to condition the battery:
1.
Insert the battery into a device that is not recording patient tests.
For details, refer to
“Replacing the Battery” on page 179
.
2.
Disconnect the AC mains power from the device.
3.
Enter the
Battery Status Service Diagnostic
window.
For details on accessing the
Battery Status Service Diagnostic
window, refer to
this device’s service manual.
4.
Allow the battery to discharge until its
Charge Level
is less than 90%.
5.
Turn off the device and reconnect the AC mains power.
6.
Allow the battery to fully charge.
The
Battery LED
is steady amber while it is charging and turns off when
charging is complete.
180
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......