Troubleshooting
Paper Jams
If the paper jams while printing, do the following:
• Verify the paper was inserted correctly.
For details, refer to
“Replacing Paper” on page 176
• Verify the paper tray spacers are set appropriately for the paper size.
For details, refer to
“Adjusting the Tray for Paper Size” on page 177
.
Import/Export/Save Errors
The following issues are discussed in this section:
•
“SD Card Not Present” on page 186
•
“Cannot Import or Transmit Records via Modem” on page 186
•
“Cannot Export to Shared Directories” on page 189
SD Card Not Present
If you receive an error message stating that the SD card is not present or cannot be
found, do the following:
• Verify an SD card is inserted into the card slot on the device.
For details, refer to
“Rear View” on page 32
.
• Verify the SD card is seated firmly.
The SD card clicks into place when seated firmly.
• Verify the SD card is formatted for a
FAT
or
FAT16
file system.
To verify an SD card is formatted for the correct file system, do the following:
1.
Insert the card into an SD card reader attached to a PC.
2.
Copy any files you want to save from the SD card to a folder on the PC.
3.
Using the Windows
Format
command, specify either
FAT
or
FAT16
for the file
system and format the card.
NOTE:
Formatting the SD card erases any existing files on the card.
4.
Copy the files from the folder on the PC to the newly formatted SD card.
Cannot Import or Transmit Records via Modem
If you receive an error while attempting to import or transmit ECG records via modem,
do the following:
• Verify the correct communication option was activated.
The system supports two options for communicating via modem:
MODC
(for
communicating with a CardioSoft/CS system) and
MODM
(for communicating with
a MUSE system). For more information, refer to
“Options Setup” on page 165
• Verify the modem is connected to an analog telephone line using a standard
RJ11 phone jack.
For more information, refer to
“Rear View” on page 32
.
186
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......