Data Input Limits
COBE
®
Spectra™ Apheresis System •
Essentials Guide
14-3
Desired TPE/RBCX fluid balance (%)
100%
75%
150%
Volume of replacement fluid to be used for TPE (mL)
(no default)
100 mL
9999 mL
Volume of plasma plus AC to remove for TPE (mL)
(no default)
100 mL
9999 mL
RBCX replacement fluid Hct (%)
80%
0%
99%
RBCX end Hct (%)
(no default)
10%
70%
Desired fraction of RBCs remaining (%)
40%
1%
100%
RBC replacement volume (mL)
0 mL
100 mL
9999 mL
Access pressure low limit (mm Hg)
–250 mm Hg
–50 mm Hg
–250 mm Hg
Return pressure high limit (mm Hg)
400 mm Hg
50 mm Hg
400 mm Hg
BMP bag volume (mL)
0
279 mL
10,000 mL
BMP Hct (%)
0%
15%
70%
* An accurate donor Hct is critical for performing a leukoreduced platelet collection procedure. Obtaining an actual Hct from the donor on the day of donation and entering
that Hct during donor data entry is recommended.
+ The operator cannot enter an inlet pump flow rate below 10 mL/min, but under certain circumstances the Spectra system can set a lower rate.
Table 14-1:
Donor/patient data entry limits (continued)
Parameter
Initial Setting (Default)
Minimum Limit
Maximum Limit
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...