Helpful Hints and Recovery Procedures
COBE
®
Spectra™ Apheresis System •
Essentials Guide
11-5
3
Clean the injection port of the sampling site coupler with alcohol,
then insert the metal spike of the return saline line into the injection
port.
4
Open the access and return saline line roller clamps:
Press
CONTINUE
.
5
The Spectra system primes the disposable tubing set according to
the standard prime procedure:
If ethylene oxide residuals are a concern, continue with Step
6
.
– or –
If ethylene oxide residuals are not a concern, continue Prime mode
according to the instructions for the apheresis procedure. You are
now finished with the Single-Pass Prime procedure.
6
When Prime mode is complete, press
PAUSE
. Leave the system in
pause for at least 15 minutes to allow ethylene oxide residuals to
diffuse from the disposable tubing set into the saline.
7
If you want to prime the set again, press
CHANGE MODE
. The
following screen appears:
Press
2
.
8
The following screen appears:
Press
YES
.
9
If you want to prime the set for a third time, repeat steps
7
and
8
.
You must attach a new saline bag and transfer pack to the access and
return saline lines for a third prime.
10
Once you complete priming the set, attach the access and return
saline lines to a full bag of 0.9% NaCl to provide fluids during the
procedure.
11
Continue according to the instructions for completing the apheresis
procedure.
Caution:
Once fluid enters the disposable tubing set, DO
For more information, see the following sections in
Chapter 11 of the Essentials Guide:
•
“Repositioning the Access Pressure Sensor Diaphragm”
•
“Repositioning the Return Pressure Sensor Diaphragm”
Open access and return saline lines.
Press CONTINUE to prime.
Prime access, return connections
(and SpectraTHERM line).
CONTINUE
Caution:
The volume capacity of the disposable tubing set
1=Load Set, 2=Prime, 3=Run, 4=Rinseback,
5=Unload Set, 6=Diagnostics.
Re-prime tubing set, (YES/NO)?
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...