Software Options and Configuration
COBE
®
Spectra™ Apheresis System •
Essentials Guide
5-7
Entering Donor/Patient Data After Prime Mode
At the end of Prime mode and before you connect the donor/patient, the
Spectra system prompts you to enter donor/patient data. The Spectra
system uses this data and default or configured parameters to customize
the following run parameters:
•
Pump flow rates
•
Centrifuge speed (2400rpm for platelet procedure and AutoPBSC)
•
Collect volume
•
Remove/replace volume
•
Run time
•
Procedure-specific parameters
You must enter donor/patient sex, height, weight, and hematocrit for
every procedure except BMP. The system may prompt you to enter
additional donor/patient data depending on the apheresis procedure you
are performing. Follow the instructions for the procedure to respond to
the screens. See Chapter 14, “Data Input Limits,” for the ranges allowed
during donor/patient data entry.
After you enter the data, the Spectra system calculates the run
parameters and displays them on the target run results screen. During
most procedures, you can modify the target run results to achieve the
outcome you need.
The following sections contain specific information about some of the
data entry screens. For more information, refer to the data entry
instructions in the procedural guides.
Height and Weight
You can enter donor/patient height in feet and inches, or in inches only.
If you enter inches only, enter 0 (zero) feet and the total height in inches.
Total Blood Volume (TBV)
The Spectra system calculates TBV based on the donor/patient sex,
height, and weight. You can use the calculated TBV or a different TBV,
if necessary.
For information on how the system calculates TBV and uses it for AC
management, see “System Operation and Anticoagulant Management”
on page 8-2.
During TPE procedures, the system uses TBV to calculate the
donor/patient plasma volume. During RBCX procedures, the system
uses TBV to calculate the donor/patient’s RBC volume.
Apheresis Procedure Dependant Messages
Apheresis procedures require entry of procedure specific data, such as
pre-platelet count for platelet procedures. See the procedural guides for
specific data entry requirements for each procedure. The system
calculates run results based on donor/patient data and values configured
for the procedure.
Target End/Run Results
After you enter donor/patient data, the systems displays target end
results or run results for the apheresis procedure. The information
displayed depends on the procedure you are performing.
Changing Entered Data During Run Mode
During Run mode, you can make the following changes:
•
Change the donor/patient data you entered.
•
View or change the target run results.
Note:
See “Configuring the Height/Weight Units” on page 5-3 to
select either standard or metric units.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...