General Alarms and Troubleshooting
12-50
COBE
®
Spectra™ Apheresis System •
Essentials Guide
PUMPS OFF TOO LONG. Centrifuge will turn off in __ seconds.
Alarm LED is red. Pumps stop.
To override the alarm, press
CONTINUE
.
This alarm occurs only during Run mode.
When this alarm occurs, the screen begins to count down 60 seconds.
For donor/patient safety, the centrifuge stops when the countdown
reaches zero, and the “PUMPS OFF TOO LONG” alarm occurs.
PUMPS OFF TOO LONG. Centrifuge turned off for donor/patient safety.
Alarm LED is red. Pumps and centrifuge stop.
To override the alarm, press
CONTINUE
.
This alarm occurs only during Run mode.
PUMPS OFF TOO LONG. Centrifuge will
turn off in ____ seconds.
CONTINUE*
Probable Cause
Action
Pumps paused for more than 9 minutes
If the pumps are to remain paused, continue a saline drip. The centrifuge will turn off after 60 seconds. To
restart the pumps when ready to continue the procedure, press
CONTINUE
.
PUMPS OFF TOO LONG! Centrifuge turned
off for patient safety.
CONTINUE*
Probable Cause
Action
Pumps paused for more than 10 minutes
If the pumps are to remain paused and the centrifuge is turned off, continue a saline drip. To restart the pumps
and centrifuge when ready to continue the procedure, press
CONTINUE
.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...