Alarm Tests
COBE
®
Spectra™ Apheresis System •
Essentials Guide
7-3
3
The following screen appears:
All procedures except Single-Needle ELP, LRS, and LRST:
Close
the white pinch clamps on the access and return lines near the
needle connections. Close the roller clamp on the access saline line.
Single-Needle ELP, LRS, and LRST procedures only:
Close the
white pinch clamp on the needle line between the access/return
needle and the “Y” manifold. Close the roller clamp on the access
saline line.
4
Press
CONTINUE
.
5
The Spectra system performs the tests in order. In each test, the
system moves valves and changes pump flow rates to create a
specific alarm condition. If the correct alarm occurs when the alarm
limit is reached, the Spectra system proceeds to the next alarm test.
If the correct alarm is not generated, the test fails and the following
screen appears:
See “ALARM TEST FAILED!” on page 12-19 for possible causes of
and actions to remedy this alarm. To restart the alarm tests, press
CONTINUE
.
Access Pressure Sensor Alarm Test
The following screen appears:
This test checks the operation of the access pressure sensor and detection
circuit. The system uses the inlet and AC pumps to create an alarm
condition.
When the sensor and detection circuit detect that the access pressure low
limit is reached, an alarm occurs. The following screen appears:
Primary Return Air Alarm Test
The Spectra system uses two different detection circuits to sense air in
the return air chamber. This test checks the operation of the primary
circuit. When the primary detection circuit detects air in the return air
chamber, an alarm occurs. The following screen appears:
1
Open the roller clamp on the access saline line.
Single-Needle ELP and LRS procedures:
Open the pinch clamp
below the access manifold.
2
Press
CONTINUE
.
Clamp access line and access saline.
Press CONTINUE to start Alarm Tests.
ALARM TEST FAILED!
Press CONTINUE to retry alarm tests.
Testing Access Pressure Low alarm.
ACCESS PRESSURE LOW!
Testing primary Return Air alarm.
AIR IN RETURN CHAMBER!
Open access saline line.
CONTINUE*
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...