Software Options and Configuration
5-10
COBE
®
Spectra™ Apheresis System •
Essentials Guide
AC Data
To display the AC data screen, press
4
. The following screen appears:
This screen displays the current AC infusion rate and the predicted
volume of AC in the collection and plasma bag(s) at the end of the Run.
Option 2: Pressure Display
Press
MENU ON/OFF
, then press
2
to select “Pressure Display.” The
following screen appears:
•
To display the current access and return pressures, press
1
.
•
To change the access and return pressure alarm limits during the
Run mode, press
2
.
Current Pressures
Use this screen to monitor the access and return pressures. (During
Single-Needle procedures, only the access pressure appears.) The
pressures appear on the bottom line and the pump flow rates appear on
the top line, allowing you to adjust the pump flow rates and observe the
changes in the access and return pressures at the same time.
Set Alarm Limits
You cannot view this screen before starting the Run mode. If you are in
Prime mode when you select this option, the following screen appears:
If you are in the Run mode when you select this option, the following
screen appears:
Note:
This screen is not available during RBCX or BMP
procedures.
AC infusion rate: ___ ml/min/liter TBV.
ml AC in bags: collect:___, plasma:__
1 = Current pressures,
2 = Set alarm limits.
Note:
You cannot adjust the return pressure high limit during
Single-Needle procedures.
___.__ ___.__ ___.__ ___.__ ___.__ _____
Access: ____, Return: ____ (mmHg)
Must be in Run before alarm limits
can be changed!
Access pressure low limit: –{250}
Return pressure high limit: 400
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...