Warning:
The following can cause fluid imbalances:
•
Administration of replacement fluids that are not at room
temperature
•
Use of improperly vented replacement fluid containers
•
Equipment malfunction
•
Improper line clamping or valve loading
•
During an RBCX procedure, use of an inadequately primed or
clotted RBC leukoreduction filter on the replacement line.
Refer to the filter manufacturer’s guidelines to ensure the
filter meets replacement fluid flow rates.
Monitor all solutions and procedures for correct fluid balance.
Warning:
The following can cause fluid imbalances:
•
Use of inadequately primed or clotted microaggregate filters
on the replacement line
•
Administration of replacement fluids that are not at room
temperature
•
Use of improperly vented replacement fluid containers
•
Equipment malfunction
•
Improper line clamping or valve loading
Monitor all solutions and procedures for correct fluid balance.
Warning:
The risks and complications of using fresh frozen
plasma (FFP) in plasma exchange are the same as those
associated with the therapeutic administration of FFP as described
by the American Association of Blood Banks and the American
Red Cross. However, using large volumes of FFP can increase the
frequency and severity of these complications.
Warning:
When using biologically-derived replacement fluids,
closely monitor the patient for reactions.
Warning:
DO NOT unload the disposable tubing set if the donor
or patient is connected to the Spectra system, or the donor or
patient may receive anticoagulant and other unwanted fluids
through the access and return needles.
Warning:
Opening the needle clamp and the saline line roller
clamp before taking a blood sample may result in an embolism. To
avoid a possible embolism, use the sampling bag to take a
sample. If you do not use the sampling bag, first prime the needle
with saline.
Warning:
Failure to flush the tubing with saline may result in
blood clotting in the tubing, which may result in clot(s) returning to
the donor.
Warning:
In most cases, TPE procedures will not significantly
alter the therapeutic effect of a patient’s medications. It is
advisable, however, to obtain a detailed drug history before each
procedure. For those drugs potentially affected by TPE
procedures, the physician should either adjust the doses or give
the medications immediately after the procedure.
Warning:
Repeated daily TPE procedures using plasma-poor
replacement fluids will increase the depletion of the patient’s
coagulation factors.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...