Introduction
2-22
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Functional Description
Separation
The Spectra system uses a centrifuge to separate whole blood into its
major components: Erythrocytes (red blood cells), leukocytes (white
blood cells, including peripheral blood stem cells), thrombocytes
(platelets), and plasma. The system draws whole blood from a donor or
patient, adds anticoagulant, pumps it through the tubing set and into
the centrifuge, and separates it into components. The system collects
specific components and returns remaining components to the
donor/patient. In TPE and RBCX procedures, patients receive
appropriate replacement fluid with the returned components.
During single-stage separation (TPE, RBCX, and WBC procedures), the
system pumps anticoagulated whole blood through an inlet tube into the
tubing set channel, while the channel rotates clockwise in the centrifuge.
The highest density component (RBCs) flows to the outer wall of the
channel. The buffy coat, which contains WBCs, and the plasma form
layers on top of the RBCs. Platelet collection and AutoPBSC procedures
use dual-stage separation, which is described in detail in the
Platelet
Collection Guide
and the
Cell Therapy Guide
.
The channel contains several outlet tubes. The tubes route components
for collection or return to the donor/patient, depending on the apheresis
procedure.
In the TPE and RBCX channels, two tubes route components out of the
channel. The tube that routes cellular components extends from inside
the channel to the outer channel wall, and the tube that routes plasma
extends to the inner wall. In the WBC channel, three tubes route
components out of the channel.
A control tube helps to establish and maintain the RBC/plasma
interface, so platelets or WBCs collect at the interface. The relationship
between the centrifuge speed and specific gravity of platelets and WBCs
determines which cells collect at the interface.
Saturated Fluidized Particle Bed Separation in LRS/LRS Turbo
Platelet Procedures
The collect pump directs platelet-rich plasma and the few remaining
WBCs from the second stage of the channel into the LRS chamber. The
fluid dynamics of the conical separation chamber, the sedimentation
velocity of the platelets and WBCs, and the flow rate through the
separation chamber result in an advancing platelet bed, which upon
saturation, traps WBCs in the lower levels of the LRS chamber. Platelets
exit to the storage bags.
Fluid Flows
Four variable-speed peristaltic pumps on the front panel (Figure 2-2)
control fluid flows. See page 2-5 for a description of the pumps.
When using the Spectra keypad to change a pump flow rate during a
Dual-Needle procedure, enter the actual pump flow rate for the current
procedure. An exception occurs during platelet collection procedures
when the High Flow configuration is set to “On.” When the system
establishes the RBC/plasma interface, it limits the inlet pump flow rate
to 45 mL/min. For more information on High Flow configuration, see
“Enabling the High Flow Protocol (ELP Procedures)” on page 5-6.
The system limits the inlet flow to 45 and then 65 mL/min while it
establishes a stable interface during the first part of all AutoPBSC
procedures.
When changing a pump flow rate during a Single-Needle procedure, you
actually change the average rate for that pump. At any time during a
procedure, the instantaneous pump flow rate for a specific pump can be
faster or slower than the average pump flow rate. The average pump flow
rate is the average of the pump flow rates during the Draw and Return
phases.
Note:
During ELP, LRS, LRST, TPE, and AutoPBSC procedures,
you can change only the inlet pump flow rate.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...