Specifications
13-10
COBE
®
Spectra™ Apheresis System •
Essentials Guide
AC Pump Alarm Responses
AC Pump Out-of-Range Alarm
AC Pump Runaway Alarm
Characteristics
Performance
Conditions
Detection method
AC pump equipped with encoder sensor to measure
pump flow rate as pump rotates. Measured pump flow
rate is compared to commanded pump flow rate. If
measured error is out of range for delay time, alarm
occurs.
Accuracy
Measured pump flow rate is less than 15% or greater
than 50% of commanded pump flow rate.
Delay time
Measured error remains outside range limits for
31 seconds.
Characteristics
Performance
Detection method
AC pump equipped with both an encoder sensor and Hall-effect sensor. If system is paused and either encoder
sensor or Hall-effect sensor detects the AC pump is rotating, alarm occurs.
Accuracy
Less than two revolutions of pump (approximately 2 mL) are allowed after system is paused.
Delay time
Less than 0.5 seconds from time second revolution is detected until alarm occurs.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...