Software Options and Configuration
COBE
®
Spectra™ Apheresis System •
Essentials Guide
5-11
•
To accept the default access pressure low limit of –250 mm Hg,
press
ENTER
,
– or –
Use the arrow keys or keypad to change the access pressure low limit
to between –50 and –250 mm Hg, then press
ENTER
. After you
press
ENTER
, the return pressure high limit entry is in brackets so
you can change it.
•
Use the arrow keys or keypad to change the return pressure high
limit to a value between 50 and 400 mm Hg, then press
ENTER
.
Option 3: CCM
The CCM uses two sources to transmit light through a molded cuvette
on the collect line. The Spectra system processes the light intensity at the
angular and central light-receiving points to display yields and
concentration for platelet procedures.
The system calibrates the CCM during Prime mode. If the saline
calibration sequence fails (which means that the CCM LEDs are not
turned on), the following screen appears:
To remove this screen, press
CLEAR
or
MENU ON/OFF
. For more
information about this screen, see “CCM not operational - this run
only.” on page 12-20.
Press
MENU ON/OFF
, then press
3
to select “CCM.”
ELP Procedures
The following screen appears:
•
The Current Yield is the current platelet yield collected.
•
The asterisk indicates whether the CCM is on or off. You may turn
off the CCM for the current procedure only.
•
To turn on the CCM, press
1
.
•
To turn off the CCM, press
2
.
•
The End Run Yield is the predicted yield based on the actual
number of platelets collected (as monitored by the CCM) and the
time remaining in Run mode.
•
Concentration is the predicted platelet concentration in the
collection bag at the end of the Run.
LRS and LRST Procedures
The following screen appears:
•
The Current Yield is the current platelet yield collected.
•
The asterisk indicates whether the CCM is on or off. You may turn
off the CCM for the current procedure only.
•
To turn on the CCM, press
1
.
•
To turn off the CCM, press
2
.
CCM not operational – this run only.
CCM: Current Yield=__.__E11 1=On*, 2=Off
End Run Yield=__.__E11 conc.=_______
CCM: Current Yield=__.__E11 1=On*, 2=Off
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...