Fluid Volumes
9-4
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Fluid Shifts
The amount of extracorporeal whole blood present in the Spectra system is
different from the amount of fluid outside the patient during the apheresis
procedure. At the beginning of the Run mode (see Figure 9-1), the system
has not drawn blood so the patient has no fluid volume deficit. As blood is
pumped into the system, saline from the channel is diverted to a waste bag.
At the end of Run divert, the patient fluid balance is decreased. In TPE and
RBCX procedures, if 100% fluid balance is selected, the patient remains at
this level throughout the procedure. In collection procedures (platelet and
WBC), fluid shifts are associated with the volume of product collected and
the amount of anticoagulant used. The Spectra system does not manage the
fluid balance during platelet and WBC procedures.
At the end of the procedure, the system returns RBCs to the patient. The
physician may choose to return only some of the RBCs in the channel to
avoid administering too much fluid. Table 9-2 shows the volume of blood
and saline the system returns at the end of each state of Rinseback mode.
Figure 9-1:
Patient fluid shifts during TPE procedures
250
200
150
100
50
0
-50
-100
-150
-200
-250
Fluid Shif
t (m
L)
After Run divert
Pre-Run
Run
Pre-Rinseback
Post-Rinseback
Pre-Run: The system primes the tubing set with saline but has not drawn blood from
the patient.
After Run divert: Saline is diverted to the waste bag before blood/saline returns to the
patient.
Run and pre-Rinseback: The fluid shift does not change. This example assumes a
fluid balance of 100% for TPE and RBCX procedures. Collect and AC volumes are not
accounted for in platelet and WBC procedures.
Post-Rinseback: The fluid shift increases as saline is used to rinse the channel.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...