General Alarms and Troubleshooting
12-30
COBE
®
Spectra™ Apheresis System •
Essentials Guide
EXCESSIVE LOAD ON CENTRIFUGE!
Alarm LED is red. Pumps and centrifuge stop.
To override the alarm for 60 seconds, press
CONTINUE
.
EXCESSIVE LOAD ON CENTRIFUGE!
Check centrifuge disposables.
CONTINUE
Probable Cause
Action
Channel or multi-lumen tubing improperly installed
Ensure the channel and multi-lumen tubing are properly installed in the centrifuge.
Internal system malfunction
Note the alarm message. Press
CONTINUE
.
Important:
Press
CONTINUE
only once, then call a qualified service technician. This alarm is often due to an
electrical failure, and pressing
CONTINUE
more than once can damage internal elements.
If the alarm recurs, discontinue the procedure. Press
CHANGE MODE
, then press
4
to select “Rinseback.” To
override the alarm during Rinseback mode, press
CONTINUE
. Notify a qualified service technician of the alarm
message.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...